Model 902P O2 / CO2 Analyzer
Model 902P O2 / CO2 Analyzer
| Weight | 11 kg |
|---|---|
| Dimensions | 18 × 16 × 11 cm |
| Carbon Dioxide Sensor Type |
Solid-state Infrared |
| Accuracy - Carbon Dioxide |
Accuracy (for 0-50%, 0-100% range): ± 1% of reading, or ± 0.2% (whichever is greater), Accuracy (for 0-10%, 0-20% range): ± 1% of reading or ± 0.1% CO2 (whichever is greater, Accuracy (for 0-2000 |
| CO2 Minimum Detection Limit |
0.1% (0-50% and 0-100% ranges), 0.01% (0-2.5%, 0-10% and 0-20% ranges), 10 ppm (0-5000ppm) |
| CO2 Calibration Controls |
Potentiometer SPAN adjustment for CO2. |
| CO2 Calibration Frequency |
Every 12 months; calibration gas recommended. |
| Expected Lifetime (oxygen) |
4-5 years |
| Oxygen Range |
0.0 to 100% |
| Oxygen Sensitivity |
0.1% (or optional 0.01%) |
| O2 Resolution |
0.1% O2 (optional .01%) |
| Analog Output |
0-1Vdc linear proportional to concentration |
| Accuracy |
: +/- 1% of reading; +/- 0.1% absolute below 10.0% |
| Drift |
less than 0.4% per year |
| Oxygen Sensor Type |
Proprietary Electrochemical |
| Response Time |
25 seconds to T95 |
| O2 Sensor Lifetime |
4-5 Years |
| O2 Sensor Body |
ABS Resin |
| Sample Connection |
Front panel fitting with luer connection for disposable filter, plastic splitter; 1/16 to 1/8 ID tubing; (optional) Swaglok inlet |
| Sample Method |
Pump (Optional) |
| Sample Pump |
(Optional) Miniature diaphragm type with ~5cc/sec flow |
| Pump Timing |
(Optional) 2-mode pump switch (timed, or continuous); (optional) cycling pump with on/off timer |
| Minimum Sample Required |
20cc |
| Sampling Assembly |
Inlet fitting |
| Oxygen Calibration Controls |
SPAN adjustment, zero adjustment |
| Instrument Calibration |
Calibrates with room air; coarse SPAN adjustment on back; fine span adjustment on side; ZERO adjustment on side (calibration gas recommended for CO2) |
| Exhaust port |
Vents to air OR tube outlet for recirculation to closed systems |
| Power Supply |
100-240V (50-60Hz) Charger included, with US or international interchangeable plug adapter (upon request) |
| Backup Power |
Optional 12V internal battery pack |
| Battery Operation |
Optional |
| Input Voltage |
100-240V, 50-60Hz |
| Battery Life |
8-12 hours before recharging |
| Measures: O2, CO2 |
O2, CO2 |
| Form Factor |
Benchtop |
| Size |
9.8W x 4H x 10L in. (249 x 102 x 254 mm) |
| Weight |
9.8 lb. (4.4 Kg) |
| Warranty |
2 years, parts and labor |
| Standards |
CE, RoHS |