MS3006 Multi-Angle Spectrophotometer
Measurement Geometry | 6 measurement angles (6 illumination sources, 1 receivers) |
Measure Angle | 45° Receiver: 45as-15°,45as15°,45as25°,45as45°,45as75°,45as110° Conform to the standards:ASTM D 2244,E 308,E 1164,E 2194, E2539,DIN 5033,5036,6174,6175-1,6175-2;ISO 7724, 11664-4 SAE J 1545 |
Application | Provide accurate and consistent color measurement for metallic, pearlescent and other complex special effect color products |
Light Source | Full spectrum LED light source with blue enhancement |
Lamp Life | 5 years, 3 million times measurements |
Spectrophotometric Mode | Concave Grating |
Sensor | 256 Image Element Double Array CMOS Image Sensor |
Wavelength Range | 400nm-700nm |
Wavelength Interval | 10nm |
Measurement Range | 0~600% |
Semiband Width | 10nm |
Measuring Aperture | Illumination size Φ23mm / Measuring spot 9X12mm (Customized: Illumination size Φ10mm / Measuring spot 6X8mm) |
Color Space | CIE LAB,XYZ,Yxy,LCh,βxy,DIN Lab99 |
Color Difference Formula | ΔE*ab,ΔE*94,ΔE*cmc(2:1),ΔE*cmc(1:1),ΔE*00, DINΔE99,ΔE DIN6175 |
Other Colorimetric Index | Flop Index, Int-Em |
Observer Angle | 2°/10° |
Illuminant | D65,A,C,D50,D55,D75, F1,F2(CWF),F3,F4,F5, F6,F7(DLF),F8 ,F9,F10(TPL5),F11(TL84),F12(TL83/U30) |
Display | Spectrogram/Values, Samples Chromaticity Values, Color Difference Values/Graph, PASS/FAIL Result, Color Offset, Color simulation, Sample effect value, Effect difference value |
Measuring Time | Approx. 1 second for one angle Approx. 6 seconds for all angles |
Repeatability | Spectral reflectance: Standard deviation within 0.08% Chromaticity value: ΔE*ab 0.03 ( When a white calibration plate is measured 30 times at 5 second intervals after white calibration) |
Reproducibility | ΔE*<0.10,avg on the gray tile of BCRA tile set ΔE*<0.25,avg on the color BCRA tile set |
Inter-instrument Error | 0.2ΔE*00(avg on reference Series II BCRA tile set) |
Effect Parameters | Sparkle Grade(SG),Diffuse coarseness(DC) and Color Variation(CV) |
Effect Measurement | 6 angles Sparkle Grade(SG),Color Variation(CV):15as-45°,15as-30°,15as-15°,15as15°,15as45°,15as80° 15d Diffuse coarseness(DC) |
Effect Repeatability | Sparkle Grade(SG) Short-term repeatability: 0.12% (10 times standard deviations) (When a color plate is measured 10 times at 10 second intervals after white calibration) Diffuse coarseness(DC) Short-term repeatability:e0.09% (10 times standard deviations) (When a color plate is measured 10 times at 10 second intervals after white calibration) |
Effect Reproducibility | Sparkle Grade(SG) Reproducibility: 1.9% (10 times standard deviations) (avg on reference Series II BCRA tile set) Diffuse coarseness(DC) Reproducibility: 1.4% (10 times standard deviations)(avg on reference Series II BCRA tile set) |
Trigger mode | Pressure sensing trigger, button trigger, software trigger |
Measuring Mode | Single measurement, average measurement (1-99), continuous measurement (1-99) |
Locating Mode | Color camera preview |
Dimension | L*W*H=195X83X128mm |
Weight | About 1Kg |
Power | lithium-ion battery, 3.7V,5000mAh, Continuous test 6000 times within 8 hours of full charge |
Display | 3.5-inch TFT color LCD, Capacitive Touch Screen |
Interface | USB, Bluetooth 5.0 |
Data Storage | 1000 pcs Standards,4000 pcs Samples |
Language | Simplified Chinese, Traditional Chinese, English |
Operating Environment | 10℃ to 50℃, humidity does not exceed 85%, no condensation |
Storage Environment | -20℃ to 50℃, humidity does not exceed 85%, no condensation |
Calibration | Built-in white board parameters, external white board, black cavity |
Calibration Interval | 4 hours,8 hours,24 hours,Startup calibration |
Standard accessories | Power Adapter, USB Cable, User Guide,PC Software(download from the official website), Calibration Board, black light trap,Protective cap, wristband |
Features: 1. Display the usage time, name and number of switches for each light source. 2. Automatic switching of light source with metamerism. 3. No warm-up, no flashing, fast and reliable evaluation of colors. 4. Low energy consumption, no heat dissipation, high luminous efficiency. 5. Configure a more complete standard light source for English and American standards. 6. The name of the light source can be changed, and it is more convenient to increase the light source. 7. Oversize volume and internal frame for easy observation of large item