Automatic Potentiometric Titrator HI932C2-02

Automatic Potentiometric Titrator HI932C2-02

Automatic Potentiometric Titrator HI932C2-02

Ứng dụng:

 Automatic Potentiometric Titrator HI932C2-02    

Technical Specifications

  • Analysis Types:

    • Standard Titration (standardization, preset pH/mV, pH/mV equivalence points)
    • Back Titration
    • Direct Reading
  • Endpoint Determination:

    • Preset mV
    • Preset pH
    • Up to 5 equivalence points for pH (1st or 2nd derivative)
    • Up to 5 equivalence points for mV (1st or 2nd derivative)
  • Burette:

    • Sizes: 5, 10, 25, and 50 mL
    • Resolution: 0.001 mL
    • Dispensing Rate: 0.3 mL to 2x burette volume per minute
    • Accuracy:
      • ±0.005 mL (5 mL burette)
      • ±0.010 mL (10 mL burette)
      • ±0.025 mL (25 mL burette)
      • ±0.050 mL (50 mL burette)
  • Stirring System:

    • Speed Range: 200 to 2500 rpm
    • Resolution: 100 rpm
  • pH:

    • Range: -2.0 to 20.0 pH; -2.00 ― 20.00 pH; -2.000 ― 20.000 pH
    • Resolution: 0.1 pH / 0.01 pH
    • Accuracy (@25ºC/77ºF): ±0.001 pH
    • Calibration: Up to 5 points (selectable from 8 standard buffers and 5 custom buffers)
    • Temperature Compensation: Automatic from -5.0 to 105.0°C; 23.0 to 221.0°F
  • mV:

    • Range: -2000.0 to 2000.0 mV
    • Resolution: 0.1 mV
    • Accuracy (@25ºC/77ºF): ±0.1 mV
    • Calibration: 1-point offset
  • ISE (Ion-Selective Electrode):

    • Range: 1 x 10⁻⁶ to 9.99 x 10¹⁰ concentration
    • Resolution: 1; 0.1; 0.01 concentration
    • Accuracy: ±0.5% (valence I); ±1% (valence II)
    • Calibration: Automatic, up to 5 points, 7 fixed standards (0.01, 0.1, 1, 10, 100, 1000, 10000 based on concentration) and 5 custom standards
  • Temperature:

    • Range: -5.0 ― 105.0°C; 23.0 ― 221.0°F; 268.2 ― 378.2K
    • Resolution: 0.1°C; 0.1°F; 0.1K
    • Accuracy (@25ºC/77ºF): ±0.1°C; ±0.2°F; ±0.1K (excluding electrode error)
  • Data Storage:

    • Methods: Up to 100 methods (standard and user-defined)
    • Data Logging: Up to 100 titration and pH/mV/ISE reports
Product Similar
Handheld pH/ORP/Temperature Meter Model: HI8314-1 Manufacturer: Hanna

0765 259 545

The HI9810432 is a professional, accurate and easy to use cordless pH meter from Hanna Instruments. - Specially designed to measure pH directly in cosmetic cream samples - Special junction prevents clogging - Accuracy of ±0.02 pH - Meter comes complete with calibration solutions, rinsing, filling, storage and batteries - Non-removable, pointed glass probe
4490000
Advanced Conductivity Benchtop Meter HI6321 is a streamlined Conductivity/Resistivity/Total Dissolved Solids (TDS)/Salinity/Temperature benchtop meter with a large touch screen display, comprised of a housing and an integrated conductivity measurement module. Compact and easy to operate, the HI6321 includes Hanna’s HI7631233 four-ring conductivity/resistivity/TDS/salinity probe. Recommended for a wide range of industrial process water applications, HI7631233 provides stable measurements over a wide measurement range and does not require frequent calibrations. An integral temperature sensor measures the process temperature and adjusts the measured conductivity to a reference temperature by applying specialized compensation algorithms: Linear: appropriate when it is assumed that the temperature coefficient of variation has the same value for all measurement temperatures. Standard: appropriate for high-purity water measurements and documented in ASTM Standard D5391-14. This setting should be used for >1Mohm.cm resistivity measurements. Natural: appropriate for natural ground, well, or surface water (or water with similar composition) in accordance with ISO7888 standard.