C406H Oxygen/Water Vapor Transmission Rate Test System

C406H Oxygen/Water Vapor Transmission Rate Test System

C406H Oxygen/Water Vapor Transmission Rate Test System
Model: C406H
Labthink - China
C406H Oxygen /Water Vapor Transmission Rate Test System can be used to test both OTR and WVTR of packaging materials in subsequence at preset temperature and humidity condition.

Ứng dụng:

C406H Oxygen/Water Vapor Transmission Rate Test System
Model: C406H
Labthink - China
Reference Standards
ASTM D3985, ASTM F1307, GB/T 19789, GB/T 31354, DIN 53380-3, JIS K7126-2-B, YBB 00082003-2015
ASTM F1249, ISO 15106-2, GB/T 26253, JIS K7129, YBB00092003-2015
Test Applications:
Films
The oxygen and water vapor transmission rate tests of various plastic films, paper-plastic composite films, co-extrusion films, aluminized films, aluminum foil composite films, glass fiber aluminum foil paper composite films and other film-like materials.
Sheets
The oxygen and water vapor transmission rate tests of PP sheet, PVC sheet, PVDC sheet, metal foil, rubber sheet, silicon sheet and other sheet materials.
Technical Specifications
Table 1: Test Parameters Note2
Parameter/Model
Model C406H
OTR Test Range
cc/(m2·day) (Standard area 50cm2)
0.02 ~ 200
cc/(m2·day) (MASK area 5cm2)
0.2 ~ 2000 (Optional)
cc/(m2·day) (MASK area 1cm2)
1 ~ 10000 (Optional)
OTR Resolution
cc/(m2·day)
0.0001
OTR Repeatability
cc/(m2·day)
0.02 or 1%, whichever is greater
WVTR Test Range
g/(m2·day) (Standard area 50cm2)
0.02 ~ 40
g/(m2·day) (MASK area 5cm2)
0.2 ~ 400 (Optional)
g/(m2·day) (MASK area 1cm2)
1 ~ 2000 (Optional)
WVTR Resolution
g/(m2·day)
0.0001
WVTR Repeatability
g/(m2·day)
0.02 or 2%, whichever is greater
Temperature range
°C
15 - 50
5 – 60 (Customized)
Temperature fluctuation
°C
±0.15
Humidity range
%RH (Within standard temperature range)
0%, 5 - 90%±2%, 100%
Extended functions
GP-01 Gas Purification Unit
Optional
DataShieldTM Note3
Optional
Computer System required by GMP
Optional
CFR21 Part11
Optional
Table 2: Technical Specifications
Test Chamber
6 test chambers
Specimen Size
4.6” x 4.6”(11.7cm×11.7cm)
Specimen Thickness
≤120 Mil(3mm)
Standard Test Area
50 cm2
Carrier Gas
99.999% high purity nitrogen, 99.5% oxygen (gas source is out of supply scope)
Carrier Gas Pressure
≥ 40.6 PSI / 280 kPa
Port Size
1/8 inch metal tubing
Power Supply
120VAC±10% 60Hz / 220VAC±10% 50Hz(either one of two)
Instrument Dimensions
23.6” H x 19.2” W x 25.9” D (60cm× 49cm× 66cm)
Net weight
220Lbs(100kg)
Product Similar
Oxygen Transmission Rate Test System Model: C206H Manufacturer: Labthink - China Contact for quote: 0765 259 545

0765 259 545

- Based on the rolling ball method and test the instant adhesive property of specimens - Testing steel balls are designed strictly according to national standards, which ensure the accuracy of test data - Angle of inclination could be adjusted freely to meet customer test requirements

0765 259 545

Test Dish Tightening Device Model: B2260 WVTR B2260 WVTR Test Dish Tightening Device is an auxiliary device to permeability testing instruments, for tightening the test dishes in water vapor transmission rate tests of films, sheets and similar materials used in food, medicine, medical devices, daily chemical, photovoltaic, electronic and many other industries.